Chapter 2 Easy Setting Guide
Measurement of Outer Diameter and Width (with Two Measuring Heads for Larger Objects)
A sheet cannot be located in the measuring area of a single measuring head if the width or
diameter of the sheet is excessively large. In that case, the sheet can be measured with
two measuring heads as shown below with the automatic zero function and the master
workpiece used. The result of output is displayed as OUT1.
L
L= (X+Y)
Set Values
Set item
Calibration
Tolerance setting
Area setting
Output
setting
AREA
Operation
Averaging
Measuring mode
Offset
Set value
AREA 1
Head 1 B-EDGE
Make settings if required.
A1: –1.0
A2: –1.0
Select the value according to the line speed.
Normal
Input the size of master workpiece.
Head 2 B-EDGE
No settings are required.
Set the upper and lower limit values.
AREA 2
OUT 1 OUT 2
Reference
Automatic Zero Function: Measure the master workpiece and make automatic zero
settings. (Page 3-6)
Area Settings (Page 4-7)
Calibration Settings (Page 4-20)
Output Settings (Page 4-28)
Tolerance Settings (Page 4-50)