Parameters of Commands and Responses
115 Sub display's screen * 1 R/W
0: R.V. value screen
1: Analog output screen
2: HIGH setting value screen
3: LOW setting value screen
4: Shift target value setting screen
5: Calculated value display screen
0
116
Tolerance tuning Tolerance
setting width*
9
**.*** 6 R/W 00.000 to 99.999 00.100
117 Calibration function SET1 ±**.*** 7 R/W -99.999 to +99.999 +00.000
118 Calibration function SET2 ±**.*** 7 R/W -99.999 to +99.999 +05.000
119 Calibration Function * 1 R/W
0: Standard
1: User setting
0
120 System parameter*
10
** 2 R/W 0 to 15 00
129 Calculation function*
11
*1R/W
0: The calculation function is not used.
1: Addition mode
2: Subtraction mode
3: 2 Heads mode
0
130 Measurement mode * 1 R/W
0: Edge control/Positioning mode
1: Outer diameter/Width measurement mode
2:
Inner diameter/Opening measurement mode
3: Glass edge mode
4: Pin position measurement mode
5: Pin interval judgment mode
6: Pin diameter judgment mode
7:
Specified edges interval measurement mode
0
131 Measurement direction * 1 R/W
0: Top
1: Bottom
0
132 Averaging*
12
** 2 R/W
0: hsp
1: 1
2: 2
3: 4
4: 8
5: 16
6: 32
7: 64
8: 128
9: 256
10: 512
11: 1024
12: 2048
13: 4096
05
133 Output mode * 1 R/W
0: NO (Normally Open)
1: NC (Normally Close)
0
134 Hold function setting * 1 R/W
0: Sample hold
1: Peak hold
2: Bottom hold
3: Peak-to-peak hold
4: Auto peak hold
5: Auto bottom hold
0
135
Auto peak hold or Auto bottom
hold Trigger level*
7
±**.*** 7 R/W -99.999 to +99.999 +00.100
136 Timing Input * 1 R/W
0: Level
1: Edge
0
137 Delay Timer * 1 R/W
0: Delay timer off
1: On delay timer
2: Off delay timer
3: One shot timer
0
138 Timer value **** 4 R/W 1 to 9999 0060
139 Hysteresis*
9
**.*** 6 R/W 0 to 99.999 00.010
140 Analog Output Scaling * 1 R/W
0: Initial state
1: Free range
0
141 Analog output Lower limit*
7
±**.*** 7 R/W -99.999 to +99.999 +00.000
142 Analog output Upper limit*
7
±**.*** 7 R/W -99.999 to +99.999 +10.000
143 External input*
13
*1R/W
0: Initial state
1: User setting
0
Data
No.
Data name
Data
format*
1
No. of
bytes
Attribute*
2
Data range Default value
Parameters of Commands and Responses
*1 In the Data format column, "±" indicates that the value can be either "+" or "-" and "*" signifies a
number from "0 to 9".
*2 Indicates that the data can only be read from (R), can only be written to (W), or can be both read
from and written to (R/W) the sensor amplifiers.
*3 The command is executed only when the data is changed from 0 to 1. To re-execute the
command, first change the setting to 0, then change it to 1. Otherwise, the command is not
executed.
If the sensor amplifier is turned OFF after the standard waveform registration is performed, the
previous state before the standard waveform registration is performed is restored. If you wish to
keep the standard waveform registration even after the power is turned OFF, set the saving the
standard waveform function to ON.
If the sensor amplifier is turned OFF after the zero shift is performed, the previous state before
the zero shift is performed is restored. If you wish to keep the zero shift state even after the
power is turned OFF, set the saving zero shift value function to ON.
144 External input 1*
13
*1R/W
0: Zero shift input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use
0
145 External input 2*
13
*1R/W
0: Reset input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use
0
146 External input 3*
13
*1R/W
0: Timing input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use
0
147 External input 4*
13
*1R/W
0: Gain input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use
4
148
Saving the standard waveform
function
*1R/W
0: OFF
1: ON
0
149 Saving zero shift value function * 1 R/W
0: OFF
1: ON
0
150
Interference prevention
function*
14
*1R/W
0: OFF
1: ON
0
151 Display Digit * 1 R/W
1: 0.001
2: 0.01
3: 0.1
4: 1
2
152 Power save function * 1 R/W
0: OFF
1: Half
3: All
0
153 Position monitor * 1 R/W
0: Initial state
1: OK/NG display
2: Red OFF
3: OFF
0
154 Display color * 1 R/W
0: GO Green
1: GO Red
2: Always Red
0
155
Edge check function Number of edges
** 2 R/W 0-99 01
156 Edge check function * 1 R/W
0: OFF
1: Setting A
2: Setting B
0
Data
No.
Data name
Data
format*
1
No. of
bytes
Attribute*
2
Data range Default value
Zobrazit stránku 15